University of Massachusetts Medical School Faculty Publications

Title

The Influence of Photon Attenuation on Tumor-to-Background and Signal-to-Noise Ratios for SPECT Imaging

UMMS Affiliation

Department of Radiology

Publication Date

2007-10-26

Document Type

Conference Proceeding

Disciplines

Analytical, Diagnostic and Therapeutic Techniques and Equipment | Bioimaging and Biomedical Optics | Biological and Chemical Physics | Nuclear | Nuclear Engineering | Radiology

Abstract

Expanding on the work of Nuyts et. al [1], Bai et. al. [2], and Bai and Shao [3], who all studied the effects of attenuation and attenuation correction on tumor-to-background ratios and signal detection, we have derived a general expression for the tumor-to-background ratio (TBR) for SPECT attenuated data that have been reconstructed with a linear, non-iterative reconstruction operator O. A special case of this is when O represents discrete filtered back-projection (FBP). The TBR of the reconstructed, uncorrected attenuated data (TBR(no-AC)) can be written as a weighted sum of the TBR of the FBP-reconstructed unattenuated data (TBR(FBP)) and the TBR of the FBP-reconstructed "difference" projection data (TBR(diff)). We evaluated the expression for TBR(no-AC) for a variety of objects and attenuation conditions. The ideal observer signal-to-noise ratio (SNR(ideal)) was also computed in projection space, in order to obtain an upper bound on signal detectability for a signal-known-exactly/background-known-exactly (SKE/BKE) detection task. The results generally show that SNR(ideal) is lower for tumors located deeper within the attenuating medium and increases for tumors nearer the edge of the object. In addition, larger values for the uniform attenuation coefficient mu lead to lower values for SNR(ideal). The TBR for FBP-reconstructed, uncorrected attenuated data can both under- and over-estimate the true TBR, depending on several properties of the attenuating medium, including the shape of the attenuator, the uniformity of the attenuator, and the degree to which the data are attenuated.

Keywords

Single photon emission computed tomography, Attenuation, Signal to noise ratio, Image reconstruction, Signal detection, Neoplasms, Attenuators, Genetic expression, Nonlinear filters, Upper bound

DOI of Published Version

10.1109/NSSMIC.2007.4436905

Source

Soares EJ, King MA, Byrne CL, Gifford HC, Lehovich A. The Influence of Photon Attenuation on Tumor-to-Background and Signal-to-Noise Ratios for SPECT Imaging. IEEE Nucl Sci Symp Conf Rec (1997). 2007;5:3609-3615. doi: 10.1109/NSSMIC.2007.4436905. PMID: 19169429; PMCID: PMC2630207. Link to article on publisher's site

Related Resources

Link to Article in PubMed

Journal/Book/Conference Title

IEEE Nuclear Science Symposium conference record. Nuclear Science Symposium

PubMed ID

19169429

Share

COinS