Introduction: In the active total hip replacement (THR) population, maintaining acetabular component stability and limiting polyethylene wear are crucial components to preventing premature implant failure. Titanium with Co/Cr/Mo fiber metal coating is among the most common materials used in cementless THR. Trabecular metal, composed of porous tantalum, has a metallic strut design resembling trabecular bone, designed to improve tissue infiltration and limit migration. This study assesses the stability and clinical outcomes of tantalum versus titanium acetabular shells using radiostereometric analysis (RSA) technology.

Methods: In this IRB approved, prospective, randomized, blinded study, 46 patients received a primary THR by a single surgeon (DCA). Each patient was randomized to receive a titanium (23) (Trilogy, Zimmer) or tantalum (23) (Modular tantalum shell, Zimmer) uncemented hemispheric cup and either a highly-crosslinked or conventional polyethylene liner. Tantalum RSA markers were implanted around the liner periphery, femur, and periacetabular bone in each patient. RSA examinations, Harris Hip, UCLA, WOMAC, SF-36 scores were obtained at 10 days, 6 months, and annually with the furthest patients evaluated through 5 years.

Results: Median translation was greater at all time points for the tantalum mesh cups except for the 3-year follow-up, however due to large standard errors, there was no significant difference between the two designs (p>0.05). These large standard errors were predominantly caused by two outliers, neither of which had clinical evidence of loosening at 5 years follow-up. Mean UCLA, WOMAC, Harris Hip, and SF-36 PCS and MCS scores improved similarly in both groups.

Conclusions: In this young THR population, both titanium and tantalum acetabular shells demonstrated excellent stability at five years follow up. Tantalum shells demonstrated slightly greater micromotion, but there was no statistically significant difference in shell migration. Outstanding clinical outcomes with statistically significant improvements in function and pain relief were observed in both groups.

Comments

Abstract of poster presented at the 2014 UMass Center for Clinical and Translational Science Research Retreat, held on May 20, 2014 at the University of Massachusetts Medical School, Worcester, Mass.

Creative Commons License

Creative Commons Attribution-Noncommercial-Share Alike 3.0 License
This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License.

Share

COinS
 
May 20th, 12:30 PM

Radiostereometric Analysis of Tantalum vs. Titanium Acetabular Shells in Young THR Patients

Introduction: In the active total hip replacement (THR) population, maintaining acetabular component stability and limiting polyethylene wear are crucial components to preventing premature implant failure. Titanium with Co/Cr/Mo fiber metal coating is among the most common materials used in cementless THR. Trabecular metal, composed of porous tantalum, has a metallic strut design resembling trabecular bone, designed to improve tissue infiltration and limit migration. This study assesses the stability and clinical outcomes of tantalum versus titanium acetabular shells using radiostereometric analysis (RSA) technology.

Methods: In this IRB approved, prospective, randomized, blinded study, 46 patients received a primary THR by a single surgeon (DCA). Each patient was randomized to receive a titanium (23) (Trilogy, Zimmer) or tantalum (23) (Modular tantalum shell, Zimmer) uncemented hemispheric cup and either a highly-crosslinked or conventional polyethylene liner. Tantalum RSA markers were implanted around the liner periphery, femur, and periacetabular bone in each patient. RSA examinations, Harris Hip, UCLA, WOMAC, SF-36 scores were obtained at 10 days, 6 months, and annually with the furthest patients evaluated through 5 years.

Results: Median translation was greater at all time points for the tantalum mesh cups except for the 3-year follow-up, however due to large standard errors, there was no significant difference between the two designs (p>0.05). These large standard errors were predominantly caused by two outliers, neither of which had clinical evidence of loosening at 5 years follow-up. Mean UCLA, WOMAC, Harris Hip, and SF-36 PCS and MCS scores improved similarly in both groups.

Conclusions: In this young THR population, both titanium and tantalum acetabular shells demonstrated excellent stability at five years follow up. Tantalum shells demonstrated slightly greater micromotion, but there was no statistically significant difference in shell migration. Outstanding clinical outcomes with statistically significant improvements in function and pain relief were observed in both groups.

 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.